U.S. warns over Philips Lumify devices after reports of 4 deaths
Key Facts
The U.S. Food and Drug Administration issued an early alert on October 9, 2026, about Philips Lumify ultrasound devices following reports linked to a system problem. The agency said the company had reported 4 deaths and 8 serious injuries as of September 29, 2026. The alert concerns a fault that could prevent imaging when the device is needed, alongside a separate issue affecting cable connection stability. Philips began sending correction notices to customers on September 24, 2026, before the public alert was issued. The FDA's recall record lists 13,206 units in commerce.
The agency said the Lumify app may fail to recognize a previously registered transducer and require the user to re-register it without warning before use at the point of care. According to Philips' support guidance, normal operation allows the transducer to be used after initial registration without having to repeat the process. The company said in that guidance: "After initial registration, you can use the transducer without re-registering it." The fault, however, prevents imaging until the re-registration unexpectedly requested by the app is completed. According to Philips, this process requires internet access through a wireless network or cellular connection.
The FDA identifies use outside hospitals as a leading risk setting because of limited wireless or cellular connectivity in those locations. Examples it cited include ambulances, aircraft and field locations, where completing the required registration before imaging may be difficult. The company's interim instructions call for avoiding use of the systems outside areas where wireless or cellular connectivity is readily available. This guidance remains in effect until a software update is available and installed on the affected devices. The safety instructions also recommend keeping backup ultrasound devices or alternative diagnostic methods available at all times to maintain care.
Alongside the registration fault, the agency's record identified an additional problem with the cable connection between the transducer and the phone or tablet used for imaging. According to Philips' product description, imaging is transmitted to a compatible device through a Universal Serial Bus or Lightning connection. The reported problem concerns the stability of the Universal Serial Bus connection between components of the portable system. This connection may become unstable or disconnect, disrupting the device's function during use, according to the recall record. The agency warned that a loss of function caused by this issue could delay patient diagnosis or treatment.
Philips introduced Lumify in 2015 as its first app-based ultrasound system. In an announcement dated October 14, 2016, the company said the system had been cleared for marketing for mobile and home use by healthcare professionals. The product's push into field care also included use in emergency and rescue operations outside hospital settings. On November 13, 2017, Philips announced that the German armed forces would adopt a customized version of Lumify for these operations. It said at the time that the system would become part of the standard equipment for German armed forces physicians and paramedics, in an example of its use in field care.
As part of the safety measures, Philips asked that the notice be circulated to all device users to ensure they receive the operating instructions. It also asked that the notice be posted near affected units so the instructions are available to staff when using the system. According to the alert issued on October 9, 2026, the company is developing a software update to address the transducer registration issue. It will send customers a notification and instructions when the update becomes available, while the network connectivity guidance remains in effect until it is installed. Separately, Philips is evaluating the unstable cable connection issue to determine an appropriate solution.